AbCap L 4FF
rProtein L Beads 4FF is an affinity chromatography medium designed for purification of classes, subclasses and fragments of immunoglobulins from biological fluids and from cell culture media. The recombinant protein L ligand is immobilized to highly cross-linked 4% agarose beads. The Characteristics of rProtein L Beads 4FF are summarized in Table 1.
Protein L is recombinantly expressed in E. coli and retains the property of binding to the antibody κ-chain without affecting the antigenic binding site of the antibody. Protein L binds well to the kappa light chain of human and mouse, and may bind specifically to certain kappa isoforms of other species. Protein L binds weakly to rabbit immunoglobulins and does not bind bovine, goat or sheep source immunoglobulins.
AbCap L 4FF is a prepacked ready to use ,column for purification of monoclonal and polyclonal antibodies. AbCap L 4FF 1ml and 5ml columns are packed with 1ml and 5ml of rprotein L beads 4FF. AbCap L 4FF can be adapted to all kinds of chromatography system, such as ÄKTA. It is easy to operate.
· Product Description

Table 1. Characteristics of rProtein L Beads 4FF

ItemDescription
Matrix SphericalHighly cross-linked 4% agarose beads
Ligandrecombinant protein L
Static Binding Capacity>15mg Mouse IgG/ml medium
Particle size45-165 μm
Maximum Pressure0.3 MPa, 3 bar
pH3-10
Storage Solution1×PBS containing 20% ethanol
Storage Temperature2-8℃
ProblemPossible CauseSolution
The flow rate of the column is very low.The sieve plate is blocked.Clean or replace the sieve plate.
Column is clogged.Cleaning in place(part 3).
Filtering the sample solution by passing them through a 0.22 μm or 0.45 μm filter.
The curve is not stable during sample purificationTiny air bubbles from buffer or sample.De-gas buffers and samples. Do not allow the column to dry.
A considerable amount of sample has been loaded, but no specific antibody of interest is detected.The concentration of antibody of interest is very low.Purify the antibody using the specific antigen coupled to a beads (i.e., PabPur Sufolink Beads, Cat. No. SA018005).
The antibody is sensitive to low-pH elution bufferNeutralize the eluted fractions with Neutralization Buffer immediately after elution.
The recovery rate gradually decreases.The sample is overloaded.Reduce the loading volume.
The column is too dirty and the binding capacity is reduced.Cleaning in place (part 3).
· Ordering Information
ProductCat. No.Size
rProtein L BeadsSA0450055 ml
SA04502525 ml
SA045100100 ml
SA045500500 ml
SA04501L1 L
SA04510L10 L
rProtein L Beads 4FFSA0330055 ml
SA03302525 ml
SA033100100 ml
SA033500500 ml
SA03301L1 L
SA03310L10 L
AbCap L 4FFSA033C111×1 ml
SA033C515×1 ml
SA033C151×5 ml
SA033C555×5 ml
SA033CS3×1 ml+1×5 ml
· Download Manual